Seaward PrimeTest 200 Instruções de Operação Página 17

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 20
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 16
PrimeTest 100 Operating Instructions
17
4 Specification
Earth Continuity
Accuracy* ± (5% + 2 digits)
Test current 200mA minimum
Test voltage 9V nominal
Insulation resistance
Accuracy ± (5% + 2 digits)
Test voltage 500V
Test current >1mA into 500k
Test current <2mA into 2k
Leakage Current
Accuracy ± (5% + 2 digits)
Test voltage 40V rms, 50Hz AC
Test current <5mA into 2k
Cord Test
Earth continuity, insulation resistance as above.
Check for Live and Neutral open circuit, short
circuit or reversed polarity.
*When used with Seaward test lead, Part Number 347A002
Factory Set Pass/Fail limits
Class I Class II Cord
Earth
Continuity
0.2 ohms N/A 0.2 ohm
Insulation
Resistance
1.0Mohm 2.0Mohm
2.0Mohm
Leakage
0.75mA 0.25mA N/A