Voltage
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 2%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.5%
by the Manufacturer
Temperature 0°C and 35°C E
3
1%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
3.29%
PrimeTest 200
Operating Instructions
21
Temperature
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.75%
by the Manufacturer
Temperature 0°C and 35°C E
3
1.4%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.83%
RCD
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 10%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
1%
by the Manufacturer
Temperature 0°C and 35°C E
3
0.41%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
11.24%