Appendix 1
Class 1 Earth Bond
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.49%
by the Manufacturer
Temperature 0°C and 35°C E
3
0.4%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
5.73%
PrimeTest 200
Operating Instructions
18
Class 1 Insulation
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.52%
by the Manufacturer
Temperature 0°C and 35°C E
3
1%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.30%
IEC Earth Bond
Intrinsic Error or Reference Conditions or Code Max Error
Influence Quantity Rated Operating Conditions Seen
Intrinsic Error Reference Condions A 5%
Position Reference Positions ±90° E
1
N/A
Supply Voltage At the Limits Specified E
2
0.77%
by the Manufacturer
Temperature 0°C and 35°C E
3
0.4%
Absorbed Current of Maximum Absorbed Current
Device on Test of the Device on Test E
4
N/A
Low-frequency Low-frequency
Magnetic Field Magnetic field 0.4 kA/m E
5
N/A
Not Used Not Used E
6
N/A
Capacitance Cmax = 1uF E
7
N/A
Waveform R.M.S. CR <= 2 E
8
N/A
Operating Error B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.00%